Portable Gas Detectors

Oxygen Indicator XP-3180

生産終了

Applications

  • For safety check before working in underground construction site in the tank, manhole and tunnel etc.
  • For safety of site workers.

Features

  • Extractive type oxygen concentration indicator.
  • Convenient for measuring oxygen concentration in containers where sampling is difficult.

 

Specifications

Model XP-3180
Gas Detected Oxygen
Detection Principle Galvanic cell
Sampling Method Extractive
Detection Range 0 to 25.0 vol%
Indication Accuracy
(Under identical
conditions)
±0.3vol%
(Compliant with JIS T 8201)
Alarm Set Value 18.0vol%
Display LCD digital (with backlight)
Digital display
Digital bar display
Alarm Indication Gas alarm: Intermittent buzzer, Red lamp flashing
Error alarm: Buzzer, Red lamp flashing, LCD display
Explosion-Proof
Structure
Ex ib IIB T3 (TIIS)
Operation
Temperature
‐10 to +40 °C
Power Source 4 x AA alkaline dry battery
Battery Life*1 Up to 100 hours
Dimensions W82 x H162 x D36 mm
Weight Approx. 450g (16oz.) (excluding batteries)
Standard
Accessories
Soft case, 4 x AA alkaline dry battery,
Gas sampling hose (1meter),
Probe with filter/moisture trap
Optional
Accessories
2-30meter gas sampling hose, Gas dilution probe,
Data download software (CD-ROM software and USB cable)

*1 Battery life time is subject to environmental and storage conditions.

Approvals

ATEX_logo  ULÏ°¸.ai  CE_logo  プリント

 

Type Approved by TIIS Explosion-proof certification

Conforms to JIS T 8201

Type Approved by Class NK No. 06T605 (Designation required)

Selectable Gases

Oxygen【O2】

Related Products

Contact Us

Downloads



Product Category

Product Guide Download

Search by Keywords

Contact Us

Contact Us

For telephone inquiries, please call your nearest New Cosmos Electric office.

List of New Cosmos Electric Locations